语言选择: 中文版 line 英文版

Digital pressure gauge with 6MPa can detect how much range and accuracy level pointer gauge?

Detailed description of the problem: The digital pressure gauge is 6 MPa, 0.05 level, and it can detect a pointer table with a large range and accuracy level. How is it calculated specifically?
The errors of digital pressure gauges and pointer pressure gauges are calculated based on the full range. The accuracy levels of digital pressure gauges are generally 0.02, 0.05, 0.1, and 0.2.
Take a digital pressure gauge of grade 0.05 and 6MPa as an example:
The allowable error is ± 6MPa × 0.05% = ± 0.003MPa.
According to "JJG52-1999 spring tube type general pressure gauge, pressure vacuum gauge and vacuum gauge" about the value transfer regulations, according to 1/4 calculation
The maximum allowable error of the calibrated test table ± 0.003MPa × 4 = ± 0.012MPa
0.012MPa ÷ 4% × 100% = 0.3MPa
0.012MPa ÷ 2.5% × 100% = 0.48MPa
0.012MPa ÷ 1.6% × 100% = 0.75MPa
From the above, 6 MPa, 0.05 grade digital pressure gauges can verify (0.3-6) MPa, 4 grade and (0.48-6) MPa, 2.5 grade, (0.75-6) MPa, 1.6 grade general pressure gauge. The accuracy of other ranges can be calculated according to the above method.

栏目导航

联系我们

CONTACT US

联系人:Henry Meng

手机:+08613693229193

电话:+086-010-21786280

邮箱:sales_lina@foxmail.com

地址: No. 18,Jianshe Road, Kaixuan street,Liangxiang,Fangshan District,Beijing China